Operating temperature range
Room temp.-30~80°C
Temp. Control Accuracy
±0.1℃
Temp. distribution accuracy
±0.3℃
Cooling capacity
Approx. 420W (361Kcal/h), at liquid temperature 15℃
Temperature control
PID control
Operation functions
Fixed operation, automatic stop operation, automatic start operation
Temperature sensor
Temperature regulatorPt thermal resistor, for overheat protection K-type thermocouple
Temperature setting, display
Digital setting, digital display
Refrigerator, refrigerant
Air cooling 300W,R404A
Device circulation capacity (50/60Hz)
Maximum flow rate 2.8/3.2l/min (jet mixing)
Heater
Stainless steel heating pipe 850W
Cooling pipe
Nickel-plated copper
External circulation port
Both the outlet and return ports have an outer diameter of 11mm
Operating ambient temperature range
5~30℃
Safety device
Self-diagnosis function, overload protection relay for the refrigeration unit, pressure detection for the refrigeration unit, anti-dry heating float switch, delay function for refrigeration unit protection, key lock function, automatic overheating protection function, leakage protection switches for over-current, overheating protector
Other functions
Water valve, condenser filter, RS485 communication function, temperature output terminal (4~20mA), refrigeration unit pressure indicator, deviation correction function, power failure compensation function
Internal dimensions (width × depth × height mm)
150×300×170
Effective Internal dimensions (width × depth × height mm)
120×140×140
External dimensions (width x depth x height mm)
395×520×880